• Russia Страна-производитель

GKa-100 PZ automatic steam sterilizer

GKa-100 PZ automatic steam sterilizer is designed for sterilization with water saturated steam under excess pressure of medical products made of metal (surgical instruments, etc.); glass (dishes, etc.); rubbers (surgical gloves, etc.); plastics; dressing and ligature suture materials; articles made of textile materials (surgical, underwear, etc.), the effect of steam on which does not change their functional properties.

Пагинация
It is used in medical and prophylactic facilities and other institutions.This model is available on request.
  • 2 year warranty

  • Reliable equipment

  • Best value for money

Sterilizers are equipped with an interface connector for connection to a computer
All elements of the sterilizer vessel, except for seals, pipelines and fittings, are made of corrosion-resistant steel
3 service programs: chamber warming up (quick start), leak test, Bowie-Dick test
Microprocessor control provides constant and most accurate control of the parameters and functions of the autoclave
The sterilizer has a filter for bacterial purification of atmospheric air entering the sterilization chamber at the stage of pressure equalization

Parameter

Value

Chamber active capacity, l

100

Power, kW, not more than

11

Weight, kg, max

221±22,1

Dimensions (W*D*H), mm

600х1130х1500 (±50)

Internal dimensions, (ØxH), mm

400х771 (±3)

Selectable temperature modes

134°С- 5 min.
121°С – 20 min.

Temperature modes set at the factory on a special order

110 - 135°С
1-180 min.

Distilled water volume,
poured at the first start of the sterilizer, l

30

Temperature maintenance accuracy, °С

+3

Maximum working pressure, mPa 

0,22

Accuracy of pressure maintenance, mPa

+0,03

MTBF NMT, cycles

1000

Time of continuous work per day, h, NMT

16

Sterilizer control method

Automatic

Power supply, V/Hz

380/50

Average service life, years

10

The number of sterilization boxes of the KF-18 type, simultaneously loaded into the chamber, are supplied on special order – 4 pcs.  (supplied on a special request)

  • Vacuuming with a water ring vacuum pump
  • Preliminary removal of air from the sterilization chamber is carried out by a combined method, including the gravitational method ("blowing") and pulsating evacuation ("evacuation") via condenser
  • Full automation of the reprocessing process
  • The warranty period for the operation of the Goods is 12 months from the date of completion of commissioning, but not more than 24 months from the date of its manufacture.