VKa-75-R-«PZ» steam sterilizer

The automatic steam sterilizer for sterilization of drug solutions VKa-75-R- "PZ" is intended for sterilization of medicinal solutions with saturated water vapor in accordance with the order of the Ministry of Health of the Russian Federation No.214 dated 16.07.97 "On quality control of medicines manufactured in pharmacy organizations (pharmacies), hermetically sealed and not sealed in glass containers with a volume of up to 1000 ml".

Пагинация

It is used in pharmacies, laboratories, for sterilization of medicinal solutions hermetically sealed and not sealed in glass containers with a volume of up to 1000 ml.

  • 2 year warranty

  • Reliable equipment

  • Best value for money

The ability to program the parameters of the sterilization cycle
Central closure of the vessel lid
Sterilizers are equipped with an interface connector for connection to a computer
All elements of the sterilizer vessel, except for seals, pipelines and fittings, are made of corrosion-resistant steel, the inner surface of the chamber is electrochemically treated
Sterilization chamber and steam generator are made of stainless steel and represent a single welded structure
The drain hole of the chamber has a filter to protect the hydraulic equipment from the ingress of debris
Microprocessor control provides constant and most accurate control of the parameters and functions of the autoclave
The sterilizer has a reusable filter for bacterial purification of the air admitted to the sterilization chamber

Parameter

Value

Chamber active capacity, l

75

Power, kW, not more than

6,5

Weight, kg, max

130±13

Dimensions (W*D*H), mm

540х686х1210 (±15)

Internal dimensions, (ØxH), mm

400х674 (±3)

Selectable temperature modes

121°С - 8 min.
121°С - 12 min.
121°С - 20 min.

Free programming mode

110-126 °С/5-45 min.

Distilled water volume, poured at the first start of the sterilizer, l

35

Sterilizer heating time, NMT, min

45

Temperature maintenance accuracy, °С

±1

Maximum working pressure, mPa 

0,11

Accuracy of pressure maintenance, mPa

±0,01

MTBF NMT, cycles

1000

Time of continuous work per day, h, NMT

16

The number of sterilization boxes of the KF-18 type, simultaneously loaded into the chamber, pcs, are supplied on special order

3

Sterilizer control method

Automatic

Power supply, V/Hz

380/50

Average service life, years

10

  • Precise control of parameters and functions of the autoclave
  • Preliminary removal of air from the sterilization chamber is carried out by the method of gravity blowing off
  • Full automation of the reprocessing process
  • The warranty period for the sterilizer is 12 months from the date of completion of commissioning, but not more than 24 months from the date of manufacture